CAUTION: Potential System Damage. If the HPLC system connected to the mass spectrometer
are not controlled by the software, then do not leave the mass spectrometer unattended
while in operation. The liquid stream from the HPLC system can flood the ion source when
the mass spectrometer goes into Standby mode.
Set Queue Options
The queue goes one-by-one through the list, acquiring each sample with the selected acquisition method. After
all of the samples have been acquired, the queue stops and the mass spectrometer goes into Standby mode. In
Standby mode, the LC pumps and some instrument voltages are turned off.
The user can change the length of time the queue runs after the last acquisition has finished, before the Analyst
®
TF software puts the mass spectrometer into Standby mode. For information about the other fields in the Queue
Options dialog, refer to the Help.
1. On the Navigation bar, click
Configure
.
2. Click
Tools > Settings > Queue Options.
System User Guide
TripleTOF
®
5600/5600+ System
RUO-IDV-05-7040-B
60 / 131
8
Operating Instructions — Batches