Figure 8-1 Queue Options Dialog
3. In the
Max. Num. Waiting Samples
field, set the maximum number of samples to a value that is greater
than the number of samples that will be submitted to the queue
4. In the
Max. Idle Time
field, type the length of time the queue will wait after acquisition is completed before
going to Standby mode. The preset value is 60 minutes.
If gas cylinders are used, then adjust this time to make sure that the gas in the cylinders is not depleted.
If an LC method is used, then before the run is started, make sure that there is enough solvent in the reservoirs
for all of the sample runs at the primary flow rate and the maximum idle time.
5. Select the
Leave Mass Spec on in Standby
check box to keep the mass spectrometer running after
analysis has been completed. This feature allows the heaters and gases to continue running, even after devices
have entered Idle state, so that the ion source and entrance to the mass spectrometer are kept free of
contaminants.
6. Select the
Fail Whole Batch in Case of Missing Vial
check box to fail the entire batch when a missing
vial is encountered. If this option is not selected, then only the current sample will fail and the queue will
continue to the next sample.
7. Select the
Fail Whole Batch if Auto Calibration Fails
check box to stop the batch if auto calibration
fails.
8. Select the
Keep Calibration Data File
check box to keep the calibration data file in a subfolder in the
Data folder of the project from which samples are being submitted.
TripleTOF
®
5600/5600+ System
System User Guide
61 / 131
RUO-IDV-05-7040-B
Operating Instructions — Batches