An XIC of the selection opens in a pane below the spectrum pane. The experiment information at the top of
the pane contains the mass range and the maximum intensity in counts per second.
Generate an XIC Using the Maximum Peak
1. Open a data file containing spectra.
2. Select a range in a spectrum.
The selection is indicated in blue.
3. Click
Explore > Extract Ions > Use Maximum.
An XIC of the maximum peak specified selection opens below the spectrum pane. The experiment information
at the top of the pane contains the mass range and the maximum intensity in counts per second.
Generate an XIC Using Base Peak Masses
1. Open a data file containing spectra.
2. In a BPC, select the peak from which to extract ions.
The selection is indicated in blue.
3. Click
Explore > Extract Ions > Use Base Peak Masses.
An XIC of the specified selection opens below the spectrum pane. The experiment information at the top of
the pane shows the mass range and the maximum intensity in counts per second.
Extract Ion by Selecting Masses
1. Open a spectrum or chromatogram.
2. Click
Explore > Extract Ions > Use Dialog.
TripleTOF
®
5600/5600+ System
System User Guide
83 / 131
RUO-IDV-05-7040-B
Operating Instructions — Analyze and Explore Data