Figure 7-2 Ion Optics Path and Parameters
Scan Type
Use
Parameter
Type
Parameter
Location
All
The ISVF parameter affects the stability of
the spray and hence the signal sensitivity.
This is the voltage applied to the needle that
sprays the sample.
Source and gas
IonSpray
Voltage
Floating
(ISVF)
1
All
The IHT parameter controls the temperature
of the NanoSpray
®
interface heater and is
only available if the NanoSpray
®
ion source
and interface are installed.
The optimal heater temperature depends on
the type of sample being analyzed and the
solvent used. If the heater temperature is
too high, the signal degrades. Typically,
heater temperatures are in the 130 °C to
180 °C range. The maximum heater
temperature that can be set is 250 °C, but
this is too high for most applications.
Source and gas
Interface
Heater
Temperature
(IHT)
1
System User Guide
TripleTOF
®
5600/5600+ System
RUO-IDV-05-7040-B
56 / 131
Operating Instructions — Acquisition Methods