• If the device uses
Ethernet
as a communication interface, then type the
IP Address
assigned to the
device by the administrator or use the corresponding
Host Name
for the address.
• If the device uses
GPIB Board
as a communication interface, then do not change the settings for the GPIB
board.
The rest of the preset values for the device are probably appropriate. Do not change them. For information
about the Configuration and Communication tabs, refer to the Help.
10. To restore the device preset values, on the Communication tab, click
Set Defaults
.
11. To save the configuration, click
OK
.
12. Repeat step
13. Click
OK
in the Create New Hardware Profile dialog.
14. To activate the hardware profile, in the Hardware Configuration Editor, click the hardware profile.
15. Click
Activate Profile
.
The check mark turns green. If a red x is shown, then there is an issue with the hardware profile activation.
For more information, refer to
Troubleshoot Hardware Profile Activation on page 41
Tip!
An active hardware profile does not have to be deactivated before another one is activated. Click an
inactive hardware profile and then click
Activate Profile
. The other profile is deactivated automatically.
16. Click
Close
.
Troubleshoot Hardware Profile Activation
If a hardware profile fails to become active, then a dialog opens indicating which device in the profile failed. A
device might fail to activate because of communications errors.
1. Read the error message generated. Depending on the message, there might be an issue with a device or how
the communication is set up.
2. Verify that the device has power and is turned on.
3. Verify that the COM port assigned to the device is correct.
4. Verify that the communication settings for the device (for example, dip switch settings) are set correctly and
match the settings on the Communication tab.
5. Turn off the device.
6. Wait 10 seconds.
7. Turn on the device.
Wait until all device power-up activities are complete before trying to activate the hardware profile again.
Some devices might require 30 seconds or more to complete the power-up activities.
TripleTOF
®
5600/5600+ System
System User Guide
41 / 131
RUO-IDV-05-7040-B
Operating Instructions — Hardware Profiles and Projects