WARNING! Electrical Shock Hazard. Use only the power cables supplied with
the system. Do not use power cables that are not properly rated for the
operation of this system.
An external line transformer is not needed for the mass spectrometer or roughing pump.
Protective Earth Conductor
The mains supply must include a correctly installed protective earth conductor. The protective earth conductor
must be installed or checked by a qualified electrician before the system is connected.
WARNING! Electrical Shock Hazard. Do not intentionally interrupt the
protective earth conductor. Any interruption of the protective earth conductor
creates an electrical shock hazard.
WARNING! Electrical Shock Hazard. Make sure that a protective earth
conductor (grounding cable) is connected between the sample loop and an
appropriate grounding point at the ion source. This supplementary grounding
will reinforce the safety configuration specified by SCIEX.
Chemical Precautions
WARNING! Ionizing Radiation Hazard, Biohazard, or Toxic Chemical Hazard.
Determine whether decontamination is required prior to cleaning or
maintenance. The customer must decontaminate the system prior to cleaning
or maintenance if radioactive materials, biological agents, or toxic chemicals
have been used with the system.
WARNING! Environmental Hazard. Do not dispose of system components in municipal
waste. Follow local regulations when disposing of components.
WARNING! Biohazard, Toxic Chemical Hazard. Connect the drain tubing to
the mass spectrometer and the source exhaust drain bottle properly, to
prevent leaks.