Clean the Front of the Orifice Plate
CAUTION: Potential System Damage. When cleaning the surface of the orifice plate, do
not remove the interface heater. Frequent removal of the interface heater can result in
damage to the interface heater. Surface cleaning of the interface heater is adequate for
routine cleaning.
CAUTION: Potential System Damage. Do not insert a wire or metal brush into the aperture
on the curtain plate, orifice plate, or interface heater to avoid damaging the aperture.
1. Dampen a lint-free wipe with water and then wipe the front of the orifice plate, including the interface heater.
2. Repeat step
3. Wait until the orifice plate is dry.
4. Inspect the orifice plate for solvent stains or lint, removing any residue with a clean, slightly damp, lint-free
wipe.
Note:
Persistent spotting or filming is an indicator of contaminated solvent.
Put the Mass Spectrometer Back in Service
1. Install the curtain plate on the mass spectrometer.
2. Install the ion source on the mass spectrometer. Refer to the ion source
Operator Guide
.
Tighten the ion source by turning the source latches down into the locking position.
3. Activate the hardware profile.
Empty the Source Exhaust Drain Bottle
WARNING! Ionizing Radiation Hazard, Biohazard, or Toxic Chemical Hazard.
Deposit hazardous materials in appropriately labeled waste containers and
dispose of them according to local regulations.
TripleTOF
®
5600/5600+ System
System User Guide
101 / 131
RUO-IDV-05-7040-B
Service and Maintenance Information