Scan Type
Use
Parameter
Type
Parameter
Location
All
The GS1 parameter controls the nebulizer
gas for both the TurboIonSpray
®
and APCI
probes. The GS1 parameter controls the
nebulizer gas for the TurboIonSpray
®
probe.
Source and gas
Ion Source
Gas 1 (GS1)
1
All
The GS2 parameter controls the heater gas
for the TurboIonSpray
®
probe. The GS2
parameter controls the heater gas for the
TurboIonSpray
®
probe and the nebulizer gas
for the APCI probe.
Source and gas
Ion Source
Gas 2 (GS2)
1
All
The TEM parameter controls the
temperature of the heater gas for the
TurboIonSpray
®
probe or the temperature
of the APCI probe.
Source and gas
Temperature
(TEM)
1
All
The CUR parameter controls the gas flow of
the Curtain Gas
TM
interface. The Curtain
Gas
TM
interface is located between the
curtain plate and the orifice. It prevents the
contamination of the ion optics.
Source and gas
Curtain Gas
(CUR)
1
All
The DP parameter controls the voltage on
the orifice, which controls the ability to
decluster ions between the orifice and the
QJet
®
ion guide. It is used to minimize the
solvent clusters that might remain on the
sample ions after they enter the vacuum
chamber, and, if required, to fragment ions.
The higher the voltage, the higher the
energy imparted to the ions. If the DP
parameter is too high, then unwanted
fragmentation might occur.
Use the preset value and optimize for the
compound.
Compound
Declustering
Potential
(DP)
1
TripleTOF
®
5600/5600+ System
System User Guide
57 / 131
RUO-IDV-05-7040-B
Operating Instructions — Acquisition Methods