Operation
Page 5-12
The Analog Ouput Screen
The Analog Output screen, shown in Figure 5-9, not only allows
each of the 20 analog outputs to be viewed, but also allows the
current settings to be changed. As with the Analog Input
Screen, the values can be displayed in either engineering or
voltage units. In addition, the units of measure may be changed
for each output by pressing the F7 key. In some cases, other
outputs with the same units will also be changed to the new
units. For example, positioning the cursor over the Ozone Temp
field by using the F2 and F3 keys and pressing F7 will cause all
temperatures to be presented in units of °F; even within in
other screens.
Figure 5-9 The Analog Ouput Diagnostic Screen
The Internal Control Outputs and User Digital I/O Bits Screens
The Internal Control Outputs, shown in Figure 5-10, displays the
current states of all 24 of the Internal Control Outputs that
are used for driving internal solenoid valves and other on-off
devices. A similar screen, the User Digital I/O Bits screen
shown in Figure 5-11, displays the states of the 24 User Digital
I/O Bits that may be used for remote control and sensing of
sequences.
For both screens, each bit may serve as either an input or an
output. If an output control is inactive, the bit will sense
changes in the I/O line. If an output is active, the associated
Содержание 2010D
Страница 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Страница 3: ......
Страница 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Страница 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Страница 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Страница 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Страница 178: ......
Страница 182: ...INSTRUMENTS 2548 Shell Road Units A D Georgetown Texas 78628 Phone 512 869 0544Fax 512 869 0993 sales sabio com...