Serial Communications
Page 7-25
_________________________________________________________________
The status selection characters can be strung together in the
{status types} field in any order. The respective categories of
data will be returned in the order in which these characters
appear in the command. If a category of status information is
requested that pertains to an option that has not been enabled,
that information will be omitted from the returned data.
The status information is always presented in pre-determined
engineering units. The following units of measure are used for
all returned status data:
Flows
Standard
Cubic
Centimeters
Per
Minute
(SCCM).
Temperatures
Degrees Centigrade (°C).
Pressures
Millimeters of Mercury (mmHg).
Gas Concentrations
Parts Per Billion (PPB).
Command Format
Command
@ GS , {AAA}, {status types}, [CHK]<CR>
Where:
{status types} =
Characters indicating what types of status should be
returned.
Responses
Good Response Examples:
{status type} = D - Dilution Status
<CR> {dilution status}, [CHK]<CR>
{status type} = O - Ozone Status
<CR> {ozone status}, [CHK]<CR>
{status type} = P
– Photometer Status
<CR> {photo status}, [CHK]<CR>
{status type} = V - Perm Oven Status (with perm oven enabled)
<CR> {perm status}, [CHK]<CR>
{status type} = G
– Gas Concentration Status
<CR> {gas status}, [CHK]<CR>
{status type} = DO - Mixed Status
<CR> {dilution status}, {ozone status}, [CHK]<CR>
{status type} = DOPG
– Mixed Status
<CR> {dilution status}, {ozone status}, {photo status},
{gas status}, [CHK]<CR>
{status type} = GDO - Mixed Status
<CR> {gas status}, {dilution status}, {ozone status}, [CHK]<CR>
Содержание 2010D
Страница 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Страница 3: ......
Страница 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Страница 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Страница 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Страница 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Страница 178: ......
Страница 182: ...INSTRUMENTS 2548 Shell Road Units A D Georgetown Texas 78628 Phone 512 869 0544Fax 512 869 0993 sales sabio com...