Operation
Page 5-11
The Analog Input Screen
The Analog Input screen shown in Figure 5-8 allows each of the
24 analog inputs to be viewed in either engineering or voltage
units. All available inputs are shown in this screen even
though some are only applicable when the corresponding options
(e.g. Photometer or Perm Oven) are installed. Inputs associated
with uninstalled options are usually floating and their values
are meaningless. The last four channels (25 - 28) are internal
voltage reference and zero channels that are used for automatic
zero/span correction.
The data presented in the screen can be viewed in either
engineering or voltage units by selecting the appropriate button
on the screen and pressing enter (or by toggling the selection
with the F6 key). If the "Diag." Button is pressed while this
screen is displayed, a special debug mode is entered that allows
the A/D scanning to be halted and a particular analog input
channel to be frozen. This function is normally not needed by
the user and is intended for factory use.
Figure 5-8 The Analog Input Diagnostic Screen
Содержание 2010D
Страница 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Страница 3: ......
Страница 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Страница 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Страница 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Страница 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Страница 178: ......
Страница 182: ...INSTRUMENTS 2548 Shell Road Units A D Georgetown Texas 78628 Phone 512 869 0544Fax 512 869 0993 sales sabio com...