Serial Communications
Page 7-27
_________________________________________________________________
Source 1 Valve
Source 2 Valve
Source 3 Valve
Source 4 Valve
Source 5 Valve
Source 6 Valve
Purge Valve
Output Valve
SSSSSS
=
The states of the instument solenoids. Each
character is a 0 or 1, where 1 indicates that the
corresponding instrument solenoid output is active.
Ozone Data Fields
{o3 temp ctl}
=
The ozone lamp temperature setpoint (°C)
{o3 temp mon}
=
The measured ozone lamp temperature (°C)
{o3 lamp ctl}
=
The ozone lamp intensity setpoint (V)
{o3 lamp curr}
=
The ozone lamp current (no units)
{o3 lamp int}
=
The measured ozone lamp intensity (no units)
{o3 conc ctl}
=
The ozone concentration setpoint (PPB)
{o3 conc mon}
=
The calculated ozone concentration (PPB)
Photometer Data Fields
{photo avg}
=
The averaged ozone measurement (PPB)
{lamp temp ctl}
=
The photometer lamp temperature setpoint (°C)
{lamp temp mon}
=
The measured photometer lamp temperature (°C)
{lamp curr ctl}
=
The photometer lamp intensity setpoint (V)
{lamp curr mon}
=
The photometer lamp current (no units)
{lamp intensity}
=
The photometer lamp intensity (no units)
{det samp}
=
The photometer detector
– sample (no units)
{det ref}
=
The photometer detector
– reference (no units)
{gas temp}
=
The sample gas temperature (°C)
{gas pres}
=
The sample gas pressure (mmHg)
{gas flow}
=
The sample gas flow (SCCM)
PPP
=
The states of the photometer pump and solenoid
valves. From left to right, these bits correspond to the
following solenoid valves:
Photometer Pump
Photometer Reference Valve
Photometer Sample Valve
Perm Oven Data Fields
{perm mfc ctl}
=
The perm MFC setpoint (SCCM)
{perm mfc mon}
=
The measured perm MFC flow (SCCM)
Содержание 2010D
Страница 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Страница 3: ......
Страница 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Страница 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Страница 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Страница 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Страница 178: ......
Страница 182: ...INSTRUMENTS 2548 Shell Road Units A D Georgetown Texas 78628 Phone 512 869 0544Fax 512 869 0993 sales sabio com...