Installation
Page 4-8
Dilution Calibrator Pneumatic Connections
A rear panel drawing of the Model
2010D
Dilution Calibrator is
shown in Figure 4-1. Teflon or stainless steel tubing should be
used for pneumatic connections. Teflon tubing is preferred for
Ozone analyzer connections.
The compressed air source (Sabio Engineering Model 1001) is
normally connected to the Diluent 1 Input Port. Connect the
compressed air source input port to the system sample manifold.
The air source must be capable of supplying a constant pressure
of at least 25 psig. An alternate diluent (must be clean and
dry) may be connected to the Diluent 2 Input Port.
Diluents must
be regulated to prevent damage to the Model
2010D
.