Serial Communications
Page 7-21
_________________________________________________________________
70
Seq Start Error
An error occurred attempting to
start the seq.
71
Bad Seq Name
The sequence name is invalid.
72
Bad Seq Point
Invalid point number.
73
No Active Seq
No sequence is active.
Stop (S)
The Stop command terminates any calibrations in progress and
places the calibrator in a quiescent, idle state.
Command Format
Command
@ S , {AAA} [CHK]<CR>
Response
Good:
<ACK>
Error:
<NAK>[ER][<CR>]
Purge (P)
The Purge command causes the calibrator to purge its source
inlet manifold for a short duration.
The default duration is
five seconds, but can be changed in the
2010D
'
s dilution
parameters setup screen. If a source valve is active when the
purge command is sent, the
2010D
will purge calibration gas
through the manifold; otherwise, the gas manifold will simply be
opened to atmosphere for the purge duration.
Command Format
Command
@ P , {AAA} [CHK]<CR>
Response
Good:
<ACK>
Error:
<NAK>[ER][<CR>]
Содержание 2010D
Страница 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Страница 3: ......
Страница 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Страница 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Страница 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Страница 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Страница 178: ......
Страница 182: ...INSTRUMENTS 2548 Shell Road Units A D Georgetown Texas 78628 Phone 512 869 0544Fax 512 869 0993 sales sabio com...