Model
2010D
Calibration Setup
Page 6-25
Sequence Control Inputs
These inputs are used for Remote
Sequence Activation (see page 6-24) or in
conjunction with the Point Control Inputs
for Remote Point Activation. Each
sequence that is to be remotely activated
should have a unique pattern of Sequence
Control Inputs.
Point Control Inputs
These inputs are used in conjunction
with the Sequence Control Inputs for
Remote Point Activation (see page 6-25).
Each point that is to be remotely
activated within a particular sequence
should have a unique pattern of Point
Control Inputs for that sequence.
Point
Control Input patterns must be unique
within each sequence, but the same pattern
may be used for different sequences.
Sequence Status Outputs
These bits are configured as outputs
which signal when a particular sequence is
active. The bit pattern that is produced
on the output depends upon what was
entered during sequence setup.
More
information about setting up Sequence
Status Outputs may be found in the section
starting on page 8.
Point Status Outputs
These bits are configured as outputs
which signal when a particular point
within a sequence is active. The bit
pattern that is produced on the output
depends upon what was entered during
sequence point setup. Pages 12 and 17
contain more information about setting up
Point Status Outputs.
When setting up the digital I/O Groups, remember that no bit may
be assigned to more than one group. To assign a bit to a group,
first move the cursor to the correct bit position by using the
arrow keys and then press "1" (to un-assign a bit, press "0").
This bit pattern information should be used when mapping the
input connections to connector pins or wires when installing the
Model
2010D
as described in Section 4, Installation.
Содержание 2010D
Страница 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Страница 3: ......
Страница 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Страница 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Страница 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Страница 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Страница 178: ......
Страница 182: ...INSTRUMENTS 2548 Shell Road Units A D Georgetown Texas 78628 Phone 512 869 0544Fax 512 869 0993 sales sabio com...