Serial Communications
Page 7-28
_________________________________________________________________
{perm temp ctl}
=
The perm oven temperature setpoint (°C)
{perm temp mon}
=
The measured perm oven temperature (°C)
MMMM
=
The states of the perm oven pump and solenoid
valves. From left to right, these bits correspond to the
following solenoid valves:
Perm Vent Valve
Perm Source Valve
Perm Pump
External Perm Valve
Gas Concentration Data Fields
{total flow mon}
=
The measured total flow (SCCM)
{num gases}
=
The number of gases following.
{prim gas name}
=
The primary gas symbol (NO2, H2S, etc.)
{prim gas conc}
=
The primary gas concentration (PPB)
{gas 2 name}
=
The 2nd gas symbol (NO2, H2S, etc.)
{gas 2 conc}
=
The 2nd concentration (PPB)
{gas N name}
=
The Nth gas symbol (NO2, H2S, etc.)
{gas N conc}
=
The Nth concentration (PPB)
Response
:
<NAK>[ER][<CR>]