Serial Communications
Page 7-24
_________________________________________________________________
If a sequence is already active with the same name as {seq
name}, the next point in the sequence will be selected. If
no sequence is active or the sequence is different, the
first point (or last point if the sequence order is
descending) of the new sequence will be started.
@ MS , {AAA} , ,{point} , [CHK]<CR>
If a sequence is already active, the new point in the same
sequence will be activated. If no sequence is active, the
command will be ignored.
@ MS , {AAA} , [CHK]<CR>
If a sequence is already active, the next point in the
sequence will be activated (if it is the last point, the
sequence will be terminated). If no sequence is active,
the command will be ignored.
Get Status (GS)
The Get Status command returns current status information about
the Model
2010D
. The status generally corresponds to information
that is presented on the
2010D
'
s status screens. The response
will depend on the {status types} field, which is a list of from
one to five characters that select the categories of status
information that should be returned. The following four status
selection characters may be used in the {status type} field to
request status information from the
2010D
:
D -
Dilution Status
:
Returns the flow rates of the diluent, ozone and
source mass flow controllers and the states of the dilution
unit solenoid valves.
O -
Ozone Status:
Returns status information pertaining to the
ozone generator, if it is installed.
P -
Photometer Status:
Returns status information pertaining to the
photometer, if it is installed.
V -
Perm Oven Status
:
Returns status information pertaining to the
perm oven, if it is installed.
G -
Gas Concentration Status:
Returns the total dilution flow rate
and the names and concentrations of all gases being
produced.
Содержание 2010D
Страница 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Страница 3: ......
Страница 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Страница 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Страница 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Страница 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Страница 178: ......
Страница 182: ...INSTRUMENTS 2548 Shell Road Units A D Georgetown Texas 78628 Phone 512 869 0544Fax 512 869 0993 sales sabio com...