Model
2010D
Calibration Setup
Page 6-14
[18] Repeat steps 15 through 17 until each calibration point for
this sequence has been entered. If there more than 10
points in this sequence, press F5 after you have entered
the first 10 points. This takes you to page 3, which
allows you to program an additional 10 calibration points.
[19] When all points have been defined, the sequence information
can be reviewed by pressing F4 and F5 to page up and page
down. Information can be edited by positioning the cursor
using the F2 and F3 key to move backward or forward between
fields and then entering the new information in the
highlighted field.
When finished, press End Sequence from any page. A message
will pop up asking if you want to save the sequence.
Responding "yes" will save the sequence information you
have entered into non-volatile memory. Answering "no" will
exit, discarding any information that has been entered.
Pressing End Sequence again will return you to the New
Sequence screen without saving or discarding information.