Installation
Page 4-15
Maximum Diluent Flow = 1.00 * 10,000 cc/min
= 10,000 cc/min
Permeation Oven Flow = 94 cc/min
94 cc/min
Minimum Dilution Ratio = ------------------------
94 cc/min + 1,906 cc/min
= 4.7 * 10-2 or 21.28:1
Note: Diluent Flow = 2,000 cc/min - 94 cc/min
= 1,906 cc/min
94 cc/min
Maximum Dilution Ratio = -------------------------
94 cc/min + 10,000 cc/min
= 9.31 * 10-3 or 107.38:1
Note: Diluent Flow is maximum
Desired SO2 Instrument Span Points: 100 - 400 PPB
Oven Output Concentration Required for 100 PPB Point:
100 PPB * 107.38 = 10,738 PPB or 10.738 PPM
Maximum Concentration That Model
2010D
Can Generate:
10.738 PPM / 21.28 = 504.61 PPB
From the above calculation, a good choice would be for the Model
2010D
Permeation Oven to generate a gas concentration of 9 PPM
to ensure that the 100 PPB point can be generated. If a
permeation tube was chosen to generate a gas concentration of 9
PPM, then the following range of concentrations could be
produced:
Low Point = 9 PPM * 9.31 * 10-3
= 83.79 PPB
Содержание 2010D
Страница 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Страница 3: ......
Страница 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Страница 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Страница 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Страница 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Страница 178: ......
Страница 182: ...INSTRUMENTS 2548 Shell Road Units A D Georgetown Texas 78628 Phone 512 869 0544Fax 512 869 0993 sales sabio com...