System Operations
Page 3-7
ALT-V
Toggle the screen between voltage and
engineering units.
Button
Description
Special External Keyboard Factory Test Keys:
---
CAUTION
---
The following special keys are normally used
only at the factory for testing purposes.
Improper use of these functions or leaving
the
2010D
is a test mode could result in a
malfunction of the unit.
ALT-Q
Quick mode.
This special test mode is used for testing
sequence timing. When in quick mode, each
minute in a sequence is equal to one
second.
ALT-H
Fake Hardware.
This special test mode is used at the
factory for software testing. When fake
hardware is selected, the normal
2010D
hardware is bypassed and simulated in
software.
ALT-W
Analog input averaging mode. This
function is used for factory testing and
allows four analog input processing
options to be selected:
No Z/S Adjust
The automatic zero and span
adjustments are disabled.
No Averaging
The rolling average that
smoothes the analog input
data is disabled.
No Avg or Z/S
Both averaging and automatic
zero/span adjustments are
disabled.
Normal
Normal mode with averaging
and zero/span adjustment.
Содержание 2010D
Страница 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Страница 3: ......
Страница 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Страница 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Страница 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Страница 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Страница 178: ......
Страница 182: ...INSTRUMENTS 2548 Shell Road Units A D Georgetown Texas 78628 Phone 512 869 0544Fax 512 869 0993 sales sabio com...