Model
2010D
Calibration Setup
Page 6-26
The Digital I/O Groups may be set up by the following procedure:
[1] Select Sequences under the main menu and press the Enter
button. A pop-up menu will appear.
[2] Select Remote Activation from the menu. A pop-up menu will
appear.
[3] Select Digital I/O Groups from the pop-up menu and press
the Enter button. The Edit Digital I/O Group Assignments
screen will appear, as shown in Figure 6-9.
Figure 6-9 The Edit Digital I/O Group Assignments Screen
[4] If a separate Abort Control Input is needed (that will be
used to terminate a sequence), initialize it by setting the
desired bit to "1". When finished, press Enter to proceed
to the next field.
[5] Identify the bits that will be used for Remote Sequence
Activation by setting the bits to "1", then press Enter to
proceed to the next field.
[6] Identify the bits that will be used for Remote Point
Activation by setting the bits to "1", then press Enter to
proceed to the next field.
[7] Identify the bits that will be used as Sequence Status
Outputs by setting the bits to "1", then press Enter to
proceed to the next field.
[8] Identify the bits that will be used as Point Status Outputs
by setting the bits to "1", then press Enter to proceed to
the next field.
Содержание 2010D
Страница 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Страница 3: ......
Страница 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Страница 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Страница 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Страница 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Страница 178: ......
Страница 182: ...INSTRUMENTS 2548 Shell Road Units A D Georgetown Texas 78628 Phone 512 869 0544Fax 512 869 0993 sales sabio com...