Introduction
Page 1-4
_________________________________________________________________
Model
2010D
Gas Dilution Calibrator Specifications
Pneumatics:
Flow ac 1.0% of full scale
Flow repeata 0.15% of full scale
Flow lin 0.5% of full scale
Diluent:
Mass flow controller range 0 - 10 SLPM (Optional 0 - 20 SLPM)
Input pressure 20 - 30 PSIG
Port 1, normally air (Optional Port 2, user defined)
Source:
Mass flow controller range 0 - 100 SCCM (Optional 0 - 20 SCCM,
0 - 50 SCCM, 0 - 200 SCCM, 0 - 500 SCCM, 0 - 1 SLPM, 0 - 2 SLPM)
Optional second source mass flow controller
Input pressure 15 - 30 PSIG
Four gas cylinder input ports (Optional two additional ports)
Source manifold purge port
Optional external permeation oven input port
Usable dilution ratio varies with flow controller selection, add
a second source flow controller for wider dynamic range
Solenoid valves for diluent, source and purge ports
Eight calibration gas output ports
Optional output solenoid valve
GPT reaction chamber meets EPA requirements
Attains 95% of Set Point in less than 60 Seconds
Wetted surfaces are composed of stainless steel, Teflon, Peek,
Halar or glass; all ports Swagelok or equivalent; Viton and Calrez
seals
Содержание 2010D
Страница 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Страница 3: ......
Страница 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Страница 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Страница 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Страница 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Страница 178: ......
Страница 182: ...INSTRUMENTS 2548 Shell Road Units A D Georgetown Texas 78628 Phone 512 869 0544Fax 512 869 0993 sales sabio com...