Internal UV Ozone Photometer
Page 9-7
Following are descriptions of the items that may be found in the
Photometer section of the Ozone Calibration Status screen:
Measured Ozone
The Measured Ozone presents the average
ozone measured by the Photometer.
Lamp Temp
The Control and Monitor values of the
Photometer lamp temperature is presented
here.
Lamp Current
This is a unit-less number, between 0
and 5, that is proportional to the
current applied to and measured from the
Photometer's UV lamp.
Detector Voltage
This is the voltage that is presented
to the 2010
D
's A/D converter after
signal
conditioning
the
detector's
output. The voltage displayed here
should never approach zero or five
Volts.
Detector Count
The detector count is the value that is
actually used in the Photometer equation
on Page 2. It is a unit-less value that
is derived by applying the Detector
Offset and Gain Parameters to the
Detector Voltage.
Sample Temperature
This is the temperature of the sample
gas in the measurement tube as measured
by the Photometer's temperature sensor.
This is used to compensate the ozone
value for temperature variations.
Sample Pressure
This is the pressure in the measurement
tube as measured by the Photometer's
pressure sensor. This is used to
compensate the ozone value for pressure
variations.
Содержание 2010D
Страница 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Страница 3: ......
Страница 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Страница 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Страница 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Страница 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Страница 178: ......
Страница 182: ...INSTRUMENTS 2548 Shell Road Units A D Georgetown Texas 78628 Phone 512 869 0544Fax 512 869 0993 sales sabio com...