Model
2010D
Calibration Setup
Page 6-12
Defining Calibration Sequences" beginning on page 6-8 and ending
on page 6-10. The steps that follow assume the user has not
chosen "Multi-gas Sequence" as the sequence type.
[14] After pressing the F5 key to move to Page 2 of the Sequence
setup, the screen shown in Figure 6-5 will appear.
Figure 6-5 New Sequence Screen (Single Gas, Page 2 of 3)
The cursor will initially be in the units field for the
primary gas concentration. By pressing Enter or the down
arrow while the cursor is in the units field, the preferred
engineering units for concentrations in this sequence may be
selected from a drop-down list (%, PPM, PPB, or PPT). All
primary gas and ozone concentrations in this sequence will be
expressed in the selected units. Pressing the Enter key or F3
will proceed to the next field.
[15] After exiting the units selection field, the cursor will be
moved to a field for entering either the Primary or Ozone
concentration for point 1, depending upon the Sequence Type
that was selected on Page 1. If Gas Dilution, Gas Phase
Titration or Permeation was selected, you will be in the
column for entering the Primary concentration. If Ozone was
selected, the column of fields for entering primary gas
concentrations will be disabled and you will be moved directly
to the column for entering the Ozone concentration.
Содержание 2010D
Страница 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Страница 3: ......
Страница 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Страница 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Страница 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Страница 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Страница 178: ......
Страница 182: ...INSTRUMENTS 2548 Shell Road Units A D Georgetown Texas 78628 Phone 512 869 0544Fax 512 869 0993 sales sabio com...