Serial Communications
Page 7-6
_________________________________________________________________
Setting Up the Remote Terminal
The following terminal mode options can be changed by selecting
"Remote Terminal" under the "Comm" menu ( Figure 7-2 ):
Figure 7-2 Remote Terminal Setup Screen
Remote Session Time-out
This selects the time period that the
Model
2010D
allows after the last
keyboard
activity before
terminating
the remote
session and disconnecting.
When
the
session is terminated, the DTR line will
be dropped, causing an attached modem to
disconnect (that is, if it is set up to
handle DTR in this way, as described
earlier).
Real-Time Update Rate
Some screens, such as the status screen
and diagnostic screens, update the real-
time data at a one second rate on the
front panel, however this may be too
frequent for remote operation at lower
baud rates. This option allows the time
between screen refreshes to the remote
terminal
to
be
increased,
without
affecting the update rate on the front
panel.
Содержание 2010D
Страница 1: ...MODEL 2010D GAS DILUTION CALIBRATOR OPERATIONS MANUAL I N S T R U M E N T S...
Страница 3: ......
Страница 20: ...System Operations Page 3 2 Figure 3 1 The Model 2010D Front Panel...
Страница 42: ...Installation Page 4 15 Figure 4 1 Rear Panel...
Страница 63: ...Operation Page 5 7 Figure 5 5 The Dilution Status Screen Figure 5 6 The Ozone Status Screen...
Страница 177: ...MODEL 2010D GAS DILUTION CALIBRATOR SCHEMATICS...
Страница 178: ......
Страница 182: ...INSTRUMENTS 2548 Shell Road Units A D Georgetown Texas 78628 Phone 512 869 0544Fax 512 869 0993 sales sabio com...