Remote Control Commands - R&S
FSV-K70
R&S
®
FSV-K70
250
Operating Manual 1176.7578.02 ─ 09
Suffix:
<n>
.
1..4
Setting parameters:
<FSKRefDevAbsResult>
numeric value
Range:
The range depends on the symbol rate and has to
be between 0.1 to 15 times the symbol rate.
*RST:
100e3
Default unit: Hz
Mode:
VSA
Manual operation:
See
CALCulate<n>:STATistics:CCDF[:STATe]
<AddEvaluation>
This command switches the calculation of the statistical distribution of magnitude,
phase or frequency values on or off.
Suffix:
<n>
.
1..4
Setting parameters:
<AddEvaluation>
ON | OFF
*RST:
OFF
Example:
CALC:STAT:CCDF ON
Switches the statistic measurements on.
Mode:
VSA
Manual operation:
See
CALCulate<n>:STATistics:MODE
<StatisticMode>
This command defines whether only the symbol points or all points are considered for
the statistical calculations.
Suffix:
<n>
.
1..4
Setting parameters:
<StatisticMode>
SONLy | INFinite
SONLy
Symbol points only
INFinite
All points are used
*RST:
SONLy
Example:
CALC1:STAT:MODE SONL
Mode:
VSA
Manual operation:
See
CALCulate subsystem