Support
R&S
®
FSV-K70
353
Operating Manual 1176.7578.02 ─ 09
●
2. Press the AMPT key.
●
3. Press the "Unit" softkey.
●
4. Press the "Y-Axis Unit" softkey.
●
5. Select dB.
Problem: The Vector I/Q result display and the Constellation I/Q result display
look different
Reason:
●
The Vector I/Q diagram shows the measurement signal after the measurement fil-
ter and synchronization.
●
The Constellation I/Q diagram shows the de-rotated constellation (i.e. for a
π
/4-
DQPSK, 4 instead of 8 points are displayed). The inter-symbol interference has
been removed.
In case the measurement filter does not remove the inter-symbol interference, the
screens show measurements that are significantly different.
Problem: The Constellation I/Q measurement result display has a different num-
ber of constellation points in the R&S FSQ-K70 and the R&S FSV-K70
Reason:
In the FSQ-K70, the Constellation I/Q measurement displays the symbol instants of the
Vector I/Q measurement. Hence, this is a rotated constellation, e.g. for a
π
/4-DQPSK,
8 points are displayed.