Remote Control Commands - R&S
FSV-K70
R&S
®
FSV-K70
276
Operating Manual 1176.7578.02 ─ 09
This command is only available with option R&S
FSV-B25, but not if R&S
FSV-B17 is
active.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
INP:EATT:STAT ON
Switches the electronic attenuator into the signal path.
Manual operation:
See
INPut:FILTer:YIG[:STATe]
<State>
If the option R&S
FSVA-B11 is installed, the YIG preselector can be bypassed.
This function is only available for R&S
FSVA instruments.
If the YIG preselector at the input of the R&S
FSVA is removed from the signal path,
you can use the maximum bandwidth for signal analysis. However, image-frequency
rejection is no longer ensured.
Note that the YIG preselector is active only on frequencies greater than 7
GHz. There-
fore, switching the YIG preselector on and off has no effect if the frequency is below
that value.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1 (0 for I/Q Analyzer, GSM, VSA measurements)
Example:
INP:FILT:YIG OFF
Deactivates the YIG preselector.
Manual operation:
See
INPut:GAIN:STATe
<State>
This command turns the preamplifier on and off. (For the exact amplification value, see
the data sheet).
With option R&S
FSV-B22, the preamplifier only has an effect below 7 GHz.
With option R&S
FSV-B24, the amplifier applies to the entire frequency range.
This command is not available when using R&S Digital I/Q Interface (R&S
FSV-B17).
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
INP:GAIN:STAT ON
Turns the preamplifier on.
Manual operation:
See
INPut Subsystem
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...