Instrument Functions for Vector Signal Analysis
R&S
®
FSV-K70
87
Operating Manual 1176.7578.02 ─ 09
Figure 3-14: Result display "Phase Error" in normal mode
Available for source types:
●
Modulation Errors
SCPI commands:
CALC:FEED 'XTIM:DDEM:ERR:MPH'
to define the required source type (see
CALC:FORM PHAS
to define the result type (see
TRAC:DATA TRACE1
to query the trace results (see
3.1.1.20
Frequency Error Absolute
Displays the error of the instantaneous frequency in Hz of the measurement signal with
respect to the reference signal as a function of symbols over time.
t
FREQ
t
FREQ
t
ERR
FREQ
REF
MEAS
_
with t=n
·
T
D
and T
D
=the duration of one sampling period at the sample rate defined by
the display points per symbol parameter (see
Note that this measurement does not consider a possible carrier frequency offset. This
has already been compensated for in the measurement signal.
This measurement is mainly of interest when using the MSK or FSK modulation, but
can also be used for the PSK/QAM modulations. However, since these modulations
can have transitions through zero in the I/Q plane, in this case you might notice uncriti-
cal spikes. This is due to the fact that the phase of zero (or a complex value close to
zero) has in fact limited significance, but still influences the result of the current fre-
quency measurement.
Measurement Result Display
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...