Instrument Functions for Vector Signal Analysis
R&S
®
FSV-K70
203
Operating Manual 1176.7578.02 ─ 09
3.4.2 ASCII File Export Format for VSA Data
The data of the file header consist of three columns, each separated by a semicolon:
parameter name; numeric value; basic unit. The data section starts with the keyword
"Trace <n>" (<n> = number of stored trace), followed by the measured data in one or
several columns (depending on measurement) which are also separated by a semico-
lon.
Table 3-8: ASCII file format for VSA trace data export
File contents
Description
Header
Type;FSV;
Instrument model
Version;1.45;
Firmware version
Date;01.Apr 2010;
Date of data set storage
Screen;A;
Instrument mode
Points per Symbol;4;
Points per symbol
x Axis Start;-13;sym;
Start value of the x axis
x Axis Stop;135;sym;
Stop value of the x axis
Ref value y axis;-10.00;dBm;
Y axis reference value
Ref value position;100;%;
Y axis reference position
Data section
Trace;1;
Trace number
Meas;Result;
Result type
Meas Signal;Magnitude;
Result display
Demodulator;Offset QPSK;
Demodulation type
ResultMode;Trace;
Result mode
x unit;sym;
Unit of the x axis
y unit;dBm;
Unit of the y axis
Trace Mode;Clear Write;
Trace mode
Values;691;
Number of measurement points
10000;-10.3;-15.7
10130;-11.5;-16.9
10360;-12.0;-17.4
...;...;
Measured values: <x value>, <y1>, <y2>; <y2> being available
only with detector AUTOPEAK and containing in this case the
smallest of the two measured values for a measurement point.
Further Information
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...