Brief Description of Vector Signal Analysis
R&S
®
FSV-K70
45
Operating Manual 1176.7578.02 ─ 09
age filter is applied. The length of this filter is automatically determined with the help of
the user settings.
The filtered power of the capture buffer is subsequently compared to an automatically
chosen threshold and the rising and falling edges of bursts are identified. With the help
of the detected edges and some further processing, it is possible to decide whether the
burst "candidates" comply with the user settings.
All bursts must have a length between ("Min Burst Length" – "Search Tolerance") and
("Max Burst Length" + "Search Tolerance") to be accepted. See
on page 159 for a more detailed
description of these parameters.
Figure 2-43: Burst Search parameters
You can influence the robustness of the burst search directly by entering the correct
minimum gap length (see
on page 159), minimum burst length and
maximum burst length. Refer to
for an illustration of the three parameters.
Demodulation Overview