Auto/On/Off
Enables or disables burst searches. If "Auto" is selected, burst search is enabled only if
"Bursted Signal" is selected in the "Signal Description" tab of the "Modulation & Signal
Description" dialog box (see
"Continuous Signal / Burst Signal"
Remote command:
[SENSe:]DDEMod:SEARch:BURSt:AUTO
Meas only if burst was found
If enabled, measurement results are only displayed (and are only averaged) if a valid
burst has been found. For measurements of burst signals that are averaged over sev-
eral sweeps, this option should be enabled so that erroneous measurements do not
affect the result of averaging.
Remote command:
[SENSe:]DDEMod:SEARch:BURSt:MODE
Auto Configuration
Configures the burst search automatically. If enabled, the
Remote command:
[SENSe:]DDEMod:SEARch:BURSt:CONFigure:AUTO
Search Tolerance ← Auto Configuration
Defines the number of symbols that may differ from the burst length without influencing
the burst detection. A search tolerance of 5, for example, with a minimum and maxi-
mum burst length of 100, will detect bursts that are 95 to 100 symbols long.
Note that due to the fact that the VSA does not have knowledge of the ramp length,
there is an uncertainty in the burst search algorithm. Thus, setting this parameter to "0"
will result in a failed burst search for most signals.
Remote command:
[SENSe:]DDEMod:SEARch:BURSt:TOLerance
Min Gap Length ← Auto Configuration
Represents the minimum distance (in symbols) between adjacent bursts. The default
value is 1 symbol in order to make sure that the burst search finds bursts that are very
close to each other. However, in case the capture buffer does not contain bursts that
are narrowly Modulation Orderd, it is recommended to increase the value. This makes
the burst search faster and also more robust for highly distorted signals.
Note that this parameter only influences the robustness of the burst search. It should
not be used to explicitly exclude certain bursts from the measurement. For example,
setting the minimum gap length to 100 symbols does not ensure that the burst search
does not find bursts that are more narrowly Modulation Orderd.
Configuring VSA Measurements
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...