Remote Control Commands - R&S
FSV-K70
R&S
®
FSV-K70
302
Operating Manual 1176.7578.02 ─ 09
Setting parameters:
<MinGapLength>
numeric value
Range:
1 to 15000
*RST:
1
Default unit: SYM
Example:
DDEM:SEAR:BURS:GLEN 3US
Mode:
VSA
Manual operation:
See
[SENSe:]DDEMod:SEARch:BURSt:LENGth:MAXimum
<MaxLength>
This command defines the maximum length of a burst. Only those bursts will be recog-
nized that fall below this length. The default unit is symbols. The value can also be
given in seconds.
Setting parameters:
<MaxLength>
numeric value
Range:
0 to 15000
*RST:
1600
Default unit: SYM
Example:
DDEM:SEAR:BURS:LENG:MAX 156 us
The maximum burst length is 156
µ
s.
Mode:
VSA
[SENSe:]DDEMod:SEARch:BURSt:LENGth[:MINimum]
<UsefulLength>
This command defines the minimum length of a burst. Only those bursts will be recog-
nized that exceed this length. The default unit is symbols. The value can also be given
in seconds.
Setting parameters:
<UsefulLength>
numeric value
Range:
10 to 15000
*RST:
98
Default unit: SYM
Example:
DDEM:SEAR:BURS:LENG 140 us
The minimum burst length is 140 us.
Mode:
VSA
[SENSe:]DDEMod:SEARch:BURSt:MODE
<MeasOnlyOnBurst>
This command sets the vector analyzer so that a measurement is performed only if a
burst is found ("BURSt"). The command is available only if the burst search is activated
beforehand using the
DDEM:SEARch:BURSt:STATe = ON
command (see
SENSe subsystem
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...