Example:
DDEM:SEAR:BURS OFF
Switches burst search off.
Mode:
VSA
[SENSe:]DDEMod:SEARch:BURSt:TOLerance
<SearchTolerance>
This command controls burst search tolerance.
Setting parameters:
<SearchTolerance>
numeric value
Range:
0 to 100000
*RST:
4
Default unit: SYM
Example:
:DDEM:SEAR:BURS:TOL 1
Sets the burst tolerance to 1
Mode:
VSA
Manual operation:
See
[SENSe:]DDEMod:SEARch:MBURst:CALC
<SelectResRangeNr>
Sets the result range to be displayed after a single sweep.
Setting parameters:
<SelectResRangeNr> numeric value
Range:
1 to 1000000
*RST:
1
Default unit: NONE
Mode:
VSA
Manual operation:
See
[SENSe:]DDEMod:SEARch:MBURst:STARt?
This command queries the start of the current result range within the capture buffer.
Return values:
<StartValue>
Symbol or time at which x-axis starts