Instrument Functions for Vector Signal Analysis
R&S
®
FSV-K70
199
Operating Manual 1176.7578.02 ─ 09
241 264 773 111 337 446 514 600 67 </Data>
</RS_VSA_KNOWN_DATA_FILE>
3.3.8 Working with Limits for Modulation Accuracy Measurements
The results of a modulation accuracy measurement can be checked for violation of
defined limits automatically. If limit check is activated and the measured values exceed
the limits, those values are indicated in red in the result summary table. If limit check is
activated and no values exceed the limits, the checked values are indicated in green.
Limits and the limit check are configured in the "Limits" dialog box that is displayed
when you press the "Config ModAcc Limits" softkey in the "Lines" menu (see
To define a limit check
1. Configure a measurement with "Modulation Accuracy" as the "Source" (in the
2. Press the LINES key on the front panel.
3. Press the "Config ModAcc Limits" softkey in the "Limits" menu.
4. In the "Current" tab, define limits that the current value should not exceed for any
or all of the result types.
Note: the limits for the current value are automatically also defined for the peak
value and vice versa. However, the limit check can be activated individually for cur-
rent or peak values.
5. Select the "Check" option for each result type to be included in the limit check.
6. If necessary, define limits and activate the limit check for the mean values of the
different result types on the "Mean" tab.
7. If necessary, activate the limit check for the peak values of the different result types
on the "Peak" tab.
8. To reset the limits to their default values, press "Set to Default".
Configuring VSA Measurements
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...