Mode:
VSA
CALCulate<n>:MARKer<m>:FUNCtion:DDEMod:STATistic:SNR?
<type>
This command queries the results of the SNR error measurement performed for digital
demodulation.
Suffix:
<n>
.
1..4
screen number
<m>
1..4
irrelevant
Query parameters:
<type>
<none>
Average SNR value of current sweep
AVG
RMS Average SNR value (over several sweeps)
RPE
Peak value of SNR (over several sweeps)
SDEV
Standard deviation of SNR values over several sweeps
PCTL
95% percentile of RMS value (over several sweeps)
PEAK
Maximum SNR over all symbols of current sweep
PAVG
Average of maximum SNR values over several sweeps
TPE
Maximum SNR over all symbols over several sweeps
PSD
Standard deviation of maximum SNR values over several
sweeps
PPCT
95% percentile of maximum RMS values over several sweeps
*RST:
PEAK
Usage:
Query only
Mode:
VSA
CALCulate<n>:MARKer<m>:LINK
<MarkerCoupling>
With this command markers between several screens can be coupled, i.e. use the
same stimulus. All screens can be linked with an X-axis scaled in symbols or time,
except those showing the capture buffer. If several capture buffer measurements are
visible, their markers are coupled, too.
CALCulate subsystem
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...