Instrument Functions for Vector Signal Analysis
R&S
®
FSV-K70
126
Operating Manual 1176.7578.02 ─ 09
offset > 0:
Start of the sweep is delayed
offset < 0:
Sweep starts earlier (pre-trigger)
Only possible for span = 0 (e.g. I/Q Analyzer mode) and gated trigger
switched off
Maximum allowed range limited by the sweep time:
pretrigger
max
= sweep time
When using the R&S Digital I/Q Interface (R&S
FSV-B17) with I/Q Ana-
lyzer mode, the maximum range is limited by the number of pretrigger
samples.
See the R&S Digital I/Q Interface(R&S
FSV-B17) description in the base
unit.
In the "External" or "IF Power" trigger mode, a common input signal is used for both
trigger and gate. Therefore, changes to the gate delay will affect the trigger delay (trig-
ger offset) as well.
Remote command:
TRIGger<n>[:SEQuence]:HOLDoff[:TIME]
Trigger Offset Unit
Toggles between symbols and seconds as the trigger offset unit.
3.2.8 Softkeys of the Meas Config Menu (R&S
FSV-K70)
The "Meas Config" menu provides functions for measurement configuration.
.....................................................................................126
Settings Overview
Displays the main settings overview that visualizes the data flow of the Vector Signal
Analyzer and summarizes the current settings. In addition, the "Settings Overview" dia-
log box provides access to the individual settings dialog boxes and allows you to
restore default values.
For details on configuring the measurement and a description of the individual dialog
boxes, see
Chapter 3.3.1, "Settings Overview"
Modulation/Signal Description
Opens the "Modulation/Signal Description" dialog box.
Softkeys and Menu Overview for Vector Signal Analysis (R&S
FSV-K70)
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...