Instrument Functions for Vector Signal Analysis
R&S
®
FSV-K70
154
Operating Manual 1176.7578.02 ─ 09
Capture Length Auto
Defines the Capture Length automatically according to the burst and pattern length set-
tings and the statistics count (see
on page 146). Thus, a minimal
Capture Length is used, which improves performance.
Remote command:
Capture Length
Defines the capture length in symbols or seconds, if not defined automatically (
).
The sample rate and the usable I/Q bandwidth are displayed for reference only.
Remote command:
[SENSe:]DDEMod:RLENgth[:VALue]
Capture Oversampling
Sets the oversampling rate, i.e the number of captured points per symbol. The default
value is 4.
This parameter affects the demodulation bandwidth. If the bandwidth is too narrow, the
signal is not displayed completely. If the bandwidth is too wide, interference from out-
side the actual signal to be measured can distort the result. Thus, for signals with a
large frequency spectrum (e.g. FSK modulated signals), a higher capture oversampling
rate may be necessary.
For an indication of the required capture oversampling value, view the "Real/Imag
(I/Q)" display of the Capture Buffer with a "Spectrum" transformation. If the complete
signal is displayed and fills the width of the display, the selected value is suitable. If the
signal is cut off, increase the oversampling rate; if it is too small, decrease the over-
sampling value.
Figure 3-26: Determining the I/Q bandwidth: Real/Imag (I/Q) display of the Capture Buffer with a Spec-
trum transformation
Chapter 2.2, "Filters and Bandwidths During Signal Process-
Remote command:
Sample Rate
Shows the current sample rate.
Configuring VSA Measurements
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...