Setting parameters:
<Normalize>
ON | OFF
*RST:
ON
Example:
SENS:DDEM:NORM ON
Turn on IQ offset compensation and amplitude droop compensa-
tion
Mode:
VSA
[SENSe:]DDEMod:PRATe
<CaptOverSampling>
This command determines the number of captured points per symbol.
Setting parameters:
<CaptOverSampling> 4, 8,16, 32
*RST:
4
Example:
DDEM:PRAT 8
Sets 8 points per symbol.
Mode:
VSA
Manual operation:
See
[SENSe:]DDEMod:PRESet:CALC
This command selects the Signal Overview from the predefined tab of the display over-
view dialog box.
Example:
SENS:DDEM:PRES:CALC
Resets the screen display to the presetting.
Usage:
Event
Mode:
VSA
[SENSe:]DDEMod:PRESet:RLEVel
This command initiates automatic setting of the RF attenuation and IF gain to the level
of the applied signal.
Note:
The following command must be synchronized to the end of the autorange proc-
ess using *WAI, *OPC or *OPC?, because otherwise the autorange process will be
stopped.
Example:
SENS:DDEM:PRES:RLEV;*WAI
Performs automatic level setting
Usage:
Event
Mode:
VSA
SENSe subsystem
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...