Instrument Functions for Vector Signal Analysis
R&S
®
FSV-K70
94
Operating Manual 1176.7578.02 ─ 09
Figure 3-18: Example for result summary with current EVM peak value marked
If you want to compare the trace values to the results of the Result Summary, make
sure to match the displayed points per symbol of the trace and of the Result Summary.
Refer to
Mean value
In the "Mean" column, the linear mean of the values that are in the "Current" column is
displayed. Note that if the values are in a logarithmic representation, e.g. the I/Q Off-
set, the linear values are averaged.
Peak value
In the "Peak" column, the maximum value that occurred during several evaluations is
displayed. Note that when the value can be positive and negative, e.g. the phase error,
the maximum absolute value (maintaining its sign) is displayed. The peak value of Rho
is handled differently, since its minimum value represents the worst case. In that case,
the minimum value is displayed.
Standard Deviation
The value for the standard deviation is calculated on the linear values and then conver-
ted to the displayed unit.
Measurement Result Display