Instrument Functions for Vector Signal Analysis
R&S
®
FSV-K70
177
Operating Manual 1176.7578.02 ─ 09
Not all measurements support this function.
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>:SYMBol
Source
You can choose which signal source is to be displayed from the following options:
Source
Description
Capture Buffer
The captured I/Q data
Meas & Ref
The measurement signal or the ideal reference signal (or both)
Symbols
The detected symbols (i.e. the detected bits)
Error Vector
The difference between the complex measurement signal and the complex reference
signal:
Modulation (measurement signal - reference signal)
For example: EVM = Mag(meas - ref)
Modulation Errors
Modulation errors due to different complex samples in the measurement and the ref-
erence signal:
Modulation (measurement signal) - Modulation (reference signal)
For example: Magnitude Error = Mag(meas) - Mag(ref)
Modulation Accu-
racy
Category for measurements that provide a summary on the modulation accuracy
(e.g. the Result Summary)
Remote command:
Result Type
Defines how the signal source is evaluated and which result is displayed. The available
result types depend on the selected source type.
For more information, see
Table 3-6: Available result types depending on source type
Source Type
Result Type
Capture Buffer
Magnitude Absolute
Real/Imag (I/Q)
Frequency Absolute
Vector I/Q
Meas & Ref Signal
Magnitude Absolute
Magnitude Relative
Phase Wrap
Phase Unwrap
Frequency Absolute
Frequency Relative
Configuring VSA Measurements
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...