Remote Control Commands - R&S
FSV-K70
R&S
®
FSV-K70
218
Operating Manual 1176.7578.02 ─ 09
CALCulate<n>:LIMit:MACCuracy:EVM:PCURrent:VALue
<LimitValue>
CALCulate<n>:LIMit:MACCuracy:EVM:PMEan:VALue
<LimitValue>
CALCulate<n>:LIMit:MACCuracy:EVM:PPEak:VALue
<LimitValue>
CALCulate<n>:LIMit:MACCuracy:EVM:RCURrent:VALue
<LimitValue>
CALCulate<n>:LIMit:MACCuracy:EVM:RMEan:VALue
<LimitValue>
CALCulate<n>:LIMit:MACCuracy:EVM:RPEak:VALue
<LimitValue>
This command defines the value for the current, peak or mean EVM (peak or RMS)
limit. Note that the limits for the current and the peak value are always kept identical.
Suffix:
<n>
.
1..4
window
Setting parameters:
<LimitValue>
numeric value
Range:
0.0 to 100
*RST:
1.5
Default unit: %
Example:
CALC2:FEED 'XTIM:DDEM:MACC'
switch on result summary in screen 2
CALC2:LIM:MACC:EVM:RPE:VAL 2
define a limit of 2%
Mode:
VSA
CALCulate<n>:LIMit:MACCuracy:FDERror:CURRent:VALue
<LimitValue>
CALCulate<n>:LIMit:MACCuracy:FDERror:MEAN:VALue
<LimitValue>
CALCulate<n>:LIMit:MACCuracy:FDERror:PEAK:VALue
<LimitValue>
This command defines the lower limit for the current, peak or mean center frequency
deviation error. Note that the limits for the current and the peak value are always kept
identical.
This command is available for FSK modulation only.
Suffix:
<n>
.
1..4
window
Setting parameters:
<LimitValue>
numeric value
Range:
0.0 to 1000000
*RST:
1 kHz
Default unit: Hz
Example:
CALC2:FEED 'XTIM:DDEM:MACC'
switch on result summary in screen 2
CALC2:LIM:MACC:FDER:PEAK:VAL 1050
define a limit of 1050 Hz
Mode:
VSA
CALCulate subsystem
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...