Instrument Functions for Vector Signal Analysis
R&S
®
FSV-K70
173
Operating Manual 1176.7578.02 ─ 09
Note:
In previous versions of the R&S
FSV-K70 application this setting was referred to
as "Use Pattern For Sync". The former setting "True" corresponds to the new setting
"Pattern". The former setting "False" corresponds to the new setting "Data".
If "Auto" mode is selected, the detected data is used.
"Data"
(Default): the detected data is used for synchronization, i.e. unknown
symbols
"Pattern"
Known symbols from a defined pattern are used for synchronization
Remote command:
[SENSe:]DDEMod:SEARch:PATTern:SYNC:AUTO
[SENSe:]DDEMod:SEARch:PATTern:SYNC[:STATe]
Fine Synchronization
In addition to the coarse synchronization used for symbol decisions, a fine synchroni-
zation is available to calculate various results from the reference signal, e.g. the EVM.
However, when the signal is known to have a poor transmission quality or has a high
noise level, false symbol decisions are more frequent, which may cause spikes in the
EVM results. To improve these calculations the reference signal can be estimated from
a smaller area that includes a known symbol sequence in the input signal. In this case,
the results for the limited reference area are more precise, at the cost of less accurate
results outside this area. Thus, the result range should be set to the length of the refer-
ence area. The reference area can be defined either using a pattern or using a known
data sequence from a Known Data file. If no predefined data sequences are available
for the signal, the detected data is used by default.
If "Auto" mode is selected and a Known Data file has been loaded and activated for
use, the known data sequences are used. Otherwise, the detected data is used.
Note:
You can define a maximum symbol error rate (SER) for the known data in refer-
ence to the analyzed data. If the SER of the known data exceeds this limit, the default
synchronization using the detected data is performed.
"Known Data"
The reference signal is defined as the data sequence from the loaded
Known Data file that most closely matches the measured data.
"Pattern"
The reference signal is estimated from the defined pattern.
"Detected
Data"
(Default) The reference signal is estimated from the detected data.
Remote command:
If SER ≤
This setting is only available if "Known Data" is selected for "Fine Synchronization".
You can define a maximum symbol error rate for the known data in reference to the
analyzed data. Thus, if a wrong file was mistakenly loaded or the file proves to be
unsuitable, it is not used for synchronization. Otherwise the results would be strongly
distorted. If the SER of the known data exceeds this limit, the default synchronization
using the detected data is performed.
Configuring VSA Measurements
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...