Remote Control Commands - R&S
FSV-K70
R&S
®
FSV-K70
217
Operating Manual 1176.7578.02 ─ 09
Query parameters:
<LimitResult>
NONE | PASS | FAIL | MARGIN
NONE
No limit check result available yet.
PASS
All values have passed the limit check.
FAIL
At least one value has exceeded the limit.
MARGIN
currently not used
*RST:
NONE
Example:
CALC2:FEED 'XTIM:DDEM:MACC'
switch on result summary in screen 2
CALC2:LIM:MACC:CFER:CURR:VAL 100 Hz
define a limit of [-100;100]
CALC2:LIM:MACC:CFER:CURR:STAT ON
switch limit check ON
INIT:IMM;*WAI
do single measurement
CALC2:LIM:MACC:CFER:CURR:RESULT?
query result
Usage:
Query only
Mode:
VSA
CALCulate<n>:LIMit:MACCuracy:CFERror:CURRent:VALue
<LimitValue>
CALCulate<n>:LIMit:MACCuracy:CFERror:MEAN:VALue
<LimitValue>
CALCulate<n>:LIMit:MACCuracy:CFERror:PEAK:VALue
<LimitValue>
This command defines the limit for the current, peak or mean center frequency error
limit. Note that the limits for the current and the peak value are always kept identical.
Suffix:
<n>
.
1..4
window
Setting parameters:
<LimitValue>
numeric value
the value x (x>0) defines the interval [-x; x]
Range:
0.0 to 1000000
*RST:
1000.0 (mean: 750.0)
Default unit: Hz
Example:
CALC2:FEED 'XTIM:DDEM:MACC'
switch on result summary in screen 2
CALC2:LIM:MACC:CFER:PEAK:VAL 100 Hz
define a limit of [-100;100]
Mode:
VSA
CALCulate subsystem
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...