Brief Description of Vector Signal Analysis
R&S
®
FSV-K70
13
Operating Manual 1176.7578.02 ─ 09
2.2.1 I/Q Bandwidth
The bandwidth of the I/Q data used as input for the vector signal analysis is filtered as
described in
Chapter 2.2, "Filters and Bandwidths During Signal Processing"
on page 11. Its flat, usable bandwidth (no considerable amplitude or phase distortion)
depends on:
●
the used sample rate, which depends on the
–
defined "Symbol Rate" (see
–
defined "Capture Oversampling" (see
●
the type of input used (digital baseband input, RF input, etc)
The sample rate of the I/Q input data is:
Sample rate = Symbol rate * Capture Oversampling
Table 2-1: Value range for sample rate and symbol rate
Model/option
Max. sample rate
Min. symbol rate
Max. symbol rate
(= max. sample rate / capt. oversam-
pling)
R&S
FSVA/FSV
without bandwidth
extension option
45 MHz
25 Hz
11 MHz
with bandwidth
extension option
R&S
FSV-B70/
R&S
FSVA-B40
128 MHz
25 Hz
32 MHz
R&S
FSVA/FSV
with (active) band-
width extension
option B160
1.28 GHz
25 Hz
restricted to 160 MHz
1307.9002K39
12.5 MHz
25 Hz
3.125 MHz
Using this sample rate, the resulting I/Q data bandwidth can be determined from the
figure "Relation between maximum usable bandwidth and sample rate (RF input)" in
the base unit description (section "Working with I/Q data)" for RF input operation or the
figure "Bandwidths depending on sample rate for active digital input" in the description
of the Digital Baseband interface (R&S
FSV option B17).
The sample rate and the usable I/Q bandwidth achieved for the current settings is dis-
played in the "I/Q Capture Settings" dialog, see
2.2.2 Bandwidth Extension Option R&S
FSV-B160
The bandwidth extension option R&S
FSV-B160 provides additional hardware, which
makes a maximum usable I/Q bandwidth of 160
MHz possible. As opposed to the
R&S
FSVA/FSV base unit, the R&S
FSV-K70 application uses a software resampler
which allows for a maximum sample rate up to 1.28
GHz.
Filters and Bandwidths During Signal Processing
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...