Instrument Functions for Vector Signal Analysis
R&S
®
FSV-K70
145
Operating Manual 1176.7578.02 ─ 09
FSK Ref Deviation
The FSK Reference Deviation sets the deviation to the reference frequency.
In case of 2FSK, it indicates the distance from the reference frequency to the positive /
negative deviation frequency and in case of 4FSK, the distance to the outer positive/
negative deviation frequency.
Select "Relative" from the dropdown menu next to the input field to set the deviation as
a multiple of the symbol rate (x*SR). If you want to set the deviation as an absolute
value in Hz, select "Absolute" from the dropdown menu.
Note that this parameter is available only in combination with FSK modulated signals.
Remote command:
CALCulate<n>:FSK:DEViation:REFerence[:VALue]
CALCulate<n>:FSK:DEViation:REFerence:RELative
Modulation Mapping
The available mapping types depend on the
and
.
For more information on the modulation mapping, refer to
●
Remote command:
[SENSe:]DDEMod:MAPPing[:VALue]
[SENSe:]DDEMod:MAPPing:CATalog?
Symbol Rate
The symbol rate also determines the I/Q bandwidth of the data recording and demodu-
lation. You can change the default rate by entering a value in Hz.
For details on the possible values see
Remote command:
Transmit filter Type
Defines the type of transmit filter
An overview of available Transmit filters is provided in
.
Remote command:
To define the name of the Transmit filter to be used.
[SENSe:]DDEMod:TFILter[:STATe]
To switch off the Transmit filter.
Load User Filter ← Transmit filter Type
Opens a file-selection dialog box to select the user-defined Transmit filter to be used.
Note:
If a user-defined Transmit filter is selected and the measurement filter is defined
on page 175), a Low-ISI measurement filter according to the
selected user filter is calculated and used.
Configuring VSA Measurements
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...