Remote Control Commands - R&S
FSV-K70
R&S
®
FSV-K70
287
Operating Manual 1176.7578.02 ─ 09
[SENSe:]DDEMod:FILTer[:STATe]
<MeasFilterState>
This command defines whether the input signal that is evaluated is filtered by the mea-
surement filter. This command has no effect on the Transmit filter.
Setting parameters:
<MeasFilterState>
ON | OFF
ON
OFF
The input signal is not filtered.
*RST:
ON
Example:
DDEM:FILT OFF
The input signal is not filtered.
Mode:
VSA
[SENSe:]DDEMod:FORMat
<Group>
This command selects the digital demodulation mode.
Setting parameters:
<Group>
MSK | PSK | QAM | QPSK | FSK | UQAM
QPSK
Quad Phase Shift Key
PSK
Phase Shift Key
MSK
Minimum Shift Key
QAM
Quadrature Amplitude Modulation
*RST:
PSK
Example:
SENS:DDEM:FORM QAM
Selects QAM modulation.
Mode:
VSA
Manual operation:
See
[SENSe:]DDEMod:FSK:NSTate
<FSKNstate>
This command defines the demodulation of the FSK modulation scheme.
SENSe subsystem
Содержание FSV-K70