Remote Control Commands - R&S
FSV-K70
R&S
®
FSV-K70
291
Operating Manual 1176.7578.02 ─ 09
Example:
SENS:DDEM:MAPP 'GSM'
Sets mapping to GSM.
Mode:
VSA
Manual operation:
See
[SENSe]:DDEMod:MFILter:ALPHa
<MeasFilterAlphaBT>
This command sets the alpha value of the measurement filter.
Setting parameters:
<MeasFilterAlphaBT> numeric value
Range:
0.05 to 1.0
*RST:
0.22
Default unit: NONE
Example:
SENS:DDEM:MFIL:ALPH 0.8
Sets alpha to 0.8
Mode:
VSA
Manual operation:
See
[SENSe:]DDEMod:MFILter:AUTO
<MeasFilterAuto>
If this command is set to "ON", the measurement filter is defined automatically depend-
ing on the Transmit filter.
Setting parameters:
<MeasFilterAuto>
ON | OFF
*RST:
ON
Example:
SENS:DDEM:MFIL:AUTO ON
Mode:
VSA
Manual operation:
See
[SENSe:]DDEMod:MFILter:NAME
<Name>
This command selects a measurement filter and automatically switches it on.
For more information on measurement filters, refer to
Chapter 2.2.6, "Measurement Fil-
Setting parameters:
<Name>
Name of the measurement filter or 'User' for a user-defined filter.
An overview of available measurement filters is provided in
.
Example:
SENS:DDEM:MFIL:NAME 'RRC'
Selects the RRC measurement filter.
Mode:
VSA
SENSe subsystem
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...