Brief Description of Vector Signal Analysis
R&S
®
FSV-K70
21
Operating Manual 1176.7578.02 ─ 09
such that the intersymbol interference is low. Hence, you will probably be able to see a
clear eye diagram and an Vector I/Q diagram with a recognizable constellation. How-
ever, a filter that has low intersymbol interference
might
lead to noise enhancement,
which is commonly undesirable for a measurement filter. In order to avoid noise
enhancement, it is recommended that you:
●
a) design your own measurement filter and upload it as a user filter
●
b) select a suitable measurement filter from the list
Transferring filter files to the R&S
FSVA/FSV
You can transfer the (
.vaf
) filter files to the R&S
FSVA/FSV using a USB memory
device.
To load a user transmit (TX) filter
1. 1. Open the "Modulation" tab of the "Modulation & Signal Description" dialog.
2. Select "Transmit filter Type":
User
.
3. Select "Load User Filter".
4. Load your
.vaf
file from the USB stick.
To load a user measurement filter
1. Open the "Measurement Filter" tab of the "Demodulation & Measurement Filter"
dialog.
2. Select "Meas Filter Type":
User
.
3. Select "Load User Filter".
4. Load your
.vaf
file from the USB stick.
2.3 Symbol Mapping
Mapping or symbol mapping means that symbol numbers are assigned to points or
transitions in the I/Q plane (e.g. PSK and QAM).
In the analyzer, the mapping is required to decode the transmitted symbols from the
sampled I/Q or frequency/time data records.
Symbol Mapping
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...