Instrument Functions for Vector Signal Analysis
R&S
®
FSV-K70
152
Operating Manual 1176.7578.02 ─ 09
Attenuation Mode
Toggles the attunuation mode. In automatic mode, the RF attenuation is automatically
set as a function of the selected reference level. This ensures that the optimum RF
attenuation is always used. It is the default setting.
In manual mode, the specified RF attenuation is used irrespective of the reference
level (see
Remote command:
RF Attenuation
For
= "Manual", this value defines the attenuation irrespective of the
reference level. If electronic attenuation is enabled (option R&S
FSV-B25 only;
= "Auto"), this setting defines the mechanical attenuation.
The mechanical attenuation can be set in 10 dB steps.
The RF attenuation can be set in 5 dB steps (with option R&S
FSV-B25: 1 dB steps).
The range is specified in the data sheet. If the defined reference level cannot be set for
the set RF attenuation, the reference level is adjusted accordingly.
Note:
Values under 10 dB can only be entered via the numeric keypad or via remote
control command in order to protect the input mixer against overload.
The RF attenuation defines the level at the input mixer according to the formula:
"level
mixer
= level
input
– RF attenuation"
The maximum mixer level allowed is -10 dBm. Mixer levels above this value may lead
to incorrect measurement results, which are indicated by the "OVLD" status display.
Remote command:
El Attenuation ON/OFF
Enables and defines the electric attenuation. The attenuation can be varied in 1 dB
steps from 0 to 30 dB. Other entries are rounded to the next lower integer value.
If the defined reference level cannot be set for the given RF attenuation, the reference
level is adjusted accordingly and the warning "Limit reached" is output.
Remote command:
I/Q Capture
The "I/Q Capture" tab contains the settings for the measured I/Q data.
Note that the maximum usable I/Q bandwidth for the R&S
FSVA/FSV40 with the order
number 1307.9002K39 is 10
MHz. Therefore the maximum symbol rate for this model
is restricted, depending on the selected capture oversampling value:
Capture oversampling
Max. symbol rate
4
3.125
MHz
8
1.5625
MHz
Configuring VSA Measurements
Содержание FSV-K70
Страница 368: ...Formulae and Abbreviations R S FSV K70 368 Operating Manual 1176 7578 02 09 Formulae ...
Страница 370: ...Formulae and Abbreviations R S FSV K70 370 Operating Manual 1176 7578 02 09 Formulae ...
Страница 371: ...Formulae and Abbreviations R S FSV K70 371 Operating Manual 1176 7578 02 09 Formulae ...
Страница 372: ...Formulae and Abbreviations R S FSV K70 372 Operating Manual 1176 7578 02 09 Formulae ...